ASAIO journal : a peer-reviewed journal of the American Society for Artificial Internal Organs
-
Randomized Controlled Trial
Clinical validation of a real-time data processing system for cardiac output and arterial pressure measurement during intraoperative biventricular pacing optimization.
Biventricular pacing (BiVP) improves cardiac output (CO) and mean arterial pressure (MAP) after cardiopulmonary bypass (CPB) in selected patients at risk for acute left heart failure after cardiac surgery. Optimization of atrioventricular delay (AVD) and interventricular delay (VVD) to maximize the hemodynamic effect of pacing requires rapid and accurate data processing. Conventional post hoc data processing (PP) is accurate but time-consuming, and infeasible in the intraoperative setting. ⋯ Interexaminer reliability coefficient values ranged from 0.997 to 0.999, indicating RTP is as reliable as PP for optimization. Real-time data processing is instantaneous and therefore is more practical in a clinical setting than the PP method. Real-time data processing is useful for guiding intraoperative BiVP optimization and merits further development.