ASAIO journal : a peer-reviewed journal of the American Society for Artificial Internal Organs
-
Observational Study
Pulse Oximetry Is Unreliable in Patients on Veno-Venous Extracorporeal Membrane Oxygenation Caused by Unrecognized Carboxyhemoglobinemia.
Continuous bedside pulse oximetry (SpO2) is universally used to monitor oxygenation for patients supported on veno-venous extracorporeal membrane oxygenation (VV-ECMO). Yet, elevated carboxyhemoglobin (COHb), a known event in VV-ECMO, diminishes the reliability of SpO2. This retrospective cohort study aims to assess the accuracy of SpO2 compared with oxyhemoglobin (SaO2) and quantify COHb levels by co-oximetry in the VV-ECMO population. ⋯ During VV-ECMO, SpO2 often overestimates SaO2 by substantial margins. This is attributable to rising COHb levels proportional to duration on VV-ECMO. In this population where adequate oxygen delivery is often marginal, clinicians should be wary of the reliability of continuous pulse oximetry to assess oxygenation.
-
Extracorporeal membrane oxygenation (ECMO) is recognized as organ support for potentially reversible acute respiratory distress syndrome (ARDS). However, limited resource during the outbreak and the coagulopathy associated with coronavirus disease 2019 (COVID-19) make the utilization of venovenous (VV) ECMO highly challenging. ⋯ High blood flow and anticoagulation at higher levels than usual practice for VV ECMO may be required because of thrombotic hematologic profile of COVID-19. Among our first 24 cases (48.8 ± 8.9 years), 17 patients were weaned from ECMO after a mean duration of 19.0 ± 10.1 days and 16 of them have been discharged from ICU.