American journal of clinical pathology
-
Am. J. Clin. Pathol. · Jan 1987
Comparative StudyMeasurement of plasma fibrin D-dimer levels with the use of a monoclonal antibody coupled to latex beads.
Recently, monoclonal antibody (DD-3B6) to fibrin D-dimer was prepared and coupled to latex beads to provide a specific test (Dimertest) for fibrinolysis. The purpose of this study was to evaluate the Dimertest assay as a clinical laboratory test for the measurement of plasma fibrin D-dimer derivatives. The Dimer-test assay specifically detected 2 micrograms/mL of purified fibrin D-dimer or fibrin D-dimer/fragment E complex added to afibrinogenemic plasma but did not detect 500 micrograms/mL of either fibrinogen fragments X, D, E, or 160 micrograms/mL cross-linked fibrinogen. ⋯ Analysis by sodium dodecyl sulfate-polyacrylamide gel electrophoresis and protein blotting with the use of monoclonal antibody DD-3B6 demonstrated a protein band with similar mobility to purified D-dimer. The measurement of plasma fibrin D-dimer derivatives by the Dimertest assay is a rapid, sensitive, and specific laboratory test for fibrinolysis. The Dimertest assay has proven to be a useful addition to the clinical laboratory and should be helpful in the diagnosis and management of patients with diseases associated with fibrinolysis.