American journal of clinical pathology
-
Am. J. Clin. Pathol. · Jan 2002
Randomized Controlled Trial Comparative Study Clinical TrialPoint-of-care testing for prothrombin time, but not activated partial thromboplastin time, correlates with laboratory methods in patients receiving aprotinin or epsilon-aminocaproic acid while undergoing cardiac surgery.
Point-of-care testing (POCT) of coagulation parameters can help optimize transfusion practice in cardiac surgery. Antifibrinolytic agents may interfere with the laboratory and/or POCT coagulation assays. This randomized controlled study compared coagulation parameters obtained from a whole blood POCT coagulation device with a typical laboratory instrument in cardiac surgery patients receiving aprotinin, epsilon-aminocaproic acid, or normal saline before undergoing cardiopulmonary bypass. ⋯ For PT, the POCT device compared favorably with the laboratory method. For aPTT, the POCT device did not compare well with the laboratory method. Treatment with antifibrinolytic agents does not interfere with determination of PT.