American journal of clinical pathology
-
Am. J. Clin. Pathol. · Jun 2007
Comparative Study Clinical TrialAccuracy of roche accu-chek inform whole blood capillary, arterial, and venous glucose values in patients receiving intensive intravenous insulin therapy after cardiac surgery.
Intravenous insulin protocols are increasingly common in the intensive care unit to maintain normoglycemia. Little is known about the accuracy of point-of-care glucometers for measuring glucose in this patient population or the impact of sample source (capillary, arterial, or venous whole blood) on the accuracy of glucometer results. We compared capillary, arterial, and venous whole blood glucose values with laboratory plasma glucose values in 20 patients after cardiac surgery. ⋯ There were no significant differences between median capillary whole blood (149 mg/dL [8.3 mmol/L]) and laboratory plasma (151 mg/dL [8.4 mmol/L]) glucose levels. The median arterial (161 mg/dL [8.9 mmol/L]) and venous (162 mg/dL [9.0 mmol/L]) whole blood glucose levels were significantly higher than the median laboratory plasma glucose level. Capillary whole blood glucose levels correlate most closely with laboratory plasma glucose levels in patients receiving intensive intravenous insulin therapy after cardiac surgery.