Platelets
-
Knowledge of platelet count and function is key to ensuring appropriate hemostatic management. We hypothesized that the novel, portable TEG®6s coagulation assessment system could evaluate the contribution of both platelet count and function to clot formation. Whole-blood samples with variable platelet counts were prepared from healthy volunteers. ⋯ MA showed significant negative relationships with abciximab concentration (p < .001). Platelet function inhibition was detected by all three assays at >68% measured by LTA and quantified in the range 68.4-82% (CK), 69.4-88% (CKH), and 69.7-76% (CRT). This demonstrates the TEG®6s analyzer can accurately evaluate platelet count and function at the site-of-care.