A randomised trial comparing real-time double-lumen endobronchial tube placement with the Disposcope(®) with conventional blind placement.
Double-lumen endobronchial tube placement is challenging. This study compared double-lumen tube placement with the Disposcope® , a wireless videostylet allowing real-time visualisation, with conventional blind placement. Patients undergoing elective thoracic surgery with normal airways requiring one-lung ventilation were randomly allocated into two groups (27 patients in each group). ⋯ In the Disposcope-assisted group, the double-lumen tube was inserted in the correct side in all patients (100.0%), whereas in the conventional group, the double-lumen tube was placed in the correct side in 25 (92.6%) patients and in the wrong side in 2 (7.4%) patients; the difference was not significant (p = 0.150). In the Disposcope-assisted group, the double-lumen tube was inserted to the optimal depth in 24 (88.9%) patients, whereas in the conventional group it was inserted to the optimal depth in one (4.0%) patient. The Disposcope increased the success rate of double-lumen tube placement, and shortened the total operation time when compared with standard placement with confirmation using fibreoptic bronchoscopy, and may replace the conventional method.