Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA
-
Knee Surg Sports Traumatol Arthrosc · Mar 2003
Comparative StudyAnalysis of initial fixation strength of press-fit fixation technique in anterior cruciate ligament reconstruction. A comparative study with titanium and bioabsorbable interference screw using porcine lower limb.
We performed a controlled laboratory study to evaluate the initial fixation strength of press-fit technique. Forty porcine lower limbs were used and divided into four groups according to the method of fixation; group 1 (press-fit+1.4 mm), in which the diameter difference between the bone plug and the femoral tunnel was 1.4 mm; group 2 (press-fit+1.4 mm, 30 degrees), in which the diameter difference was the same with group 1, but the tensile loading axis was 30 degrees away from the long axis of the femoral tunnel; group 3 (titanium), in which a titanium interference screw was used for fixation; group 4 (bioabsorbable), in which a bioabsorbable interference screw was used for fixation. The graft in the press-fit group was harvested with a hollow oscillating saw with inner diameter of 9.4 mm to obtain consistent and completely circular shape of the bone plug. ⋯ The average ultimate failure load of group 1 and group 2 were not significantly different from those of group 3 and group 4. The press-fit groups demonstrated sufficient fixation strength for the rehabilitation and interference screw groups. The completely circular shape of the bone plug and increased diameter difference between the bone plug and the femoral tunnel seemed to contribute to the strong fixation.