Primary stability of four different implants for opening wedge high tibial osteotomy.
Since a significant number of implant failures have been reported in association with the procedure of open wedge valgus high tibial osteotomy, the initial biomechanical stability of different fixation devices was investigated in this study. Fifteen third generation Sawbones composite tibiae were used as a model. Four different plates were tested: a short spacer plate (OWO) (n = 4), a short spacer plate with multi-directional locking bolts (MSO) (n = 5), a prototype version of a long spacer plate with multi-directional locking bolts (MSOnew) (n = 2), and a long medial tibia plate fixator with locking bolts (MPF) (n = 4). ⋯ The highest residual stability after failure of the lateral cortex was observed in MPF as well. The results suggest that the implant design strongly influences the primary stability of medial opening wedge tibial osteotomy. A rigid long plate fixator with angle-stable locking bolts yields the best results.