Clinical chemistry
-
We evaluated a new analyzer designed for point-of-care testing of blood gases, sodium, potassium, ionized calcium, and hematocrit. The Gem Premier (Mallinckrodt) system has two components: the analyzer and a disposable cartridge. Analysis takes place in the cartridge, which contains the electrochemical sensors, the calibrants, the reagents, the sampling stylus, and the waste container. ⋯ With aqueous control materials, total imprecision (CV) was: pH, 0.10-0.18%; PCO2, 3.16-5.78%; PO2, 2.92-4.85%; sodium, 0.82-1.44%; potassium, 1.35-1.48%; ionized calcium, 0.75-1.45%; and hematocrit, 1.13-1.83%. Accuracy of the system was assessed by split-sample comparison with the Radiometer ABL 330 blood gas analyzer for pH and blood gases, the Nova Stat Profile 5 for whole-blood electrolyte and hematocrit analysis, and the IL Phoenix for plasma electrolyte analysis. After outlier correction, regression statistics were excellent for all analytes except sodium, which demonstrated Sy[x values between 1.80 and 2.30 mmol/L and 0.85 < or = r < 0.90.