Clinical chemistry
-
Comparative Study
Impact of point-of-care testing on patients' length of stay in a large emergency department.
We prospectively investigated whether routine use of a point-of-care testing (POCT) device by nonlaboratory operators in the emergency department (ED) for all patients requiring the available tests could shorten patient length of stay (LOS) in the ED. ED patient LOS, defined as the length of time between triage (initial patient interview) and discharge (released to home or admitted to hospital), was examined during a 5-week experimental period in which ED personnel used a hand-held POCT device to perform Na, K, Cl, glucose (Gluc), and blood urea nitrogen (BUN) testing. Preliminary data demonstrated acceptable accuracy of the hand-held device. ⋯ Median LOS during the experimental period was 209 min vs 201 min for the combined control periods. Stratifying patients by presenting condition (chest pain, trauma, etc.), discharge/admit status, or presence/absence of other central laboratory tests did not reveal a decrease in patient LOS for any patient subgroup during the experimental period. From these observations, we consider it unlikely that routine use of a hand-held POCT device in a large ED such as ours is sufficient by itself to impact ED patient LOS.