Clinical chemistry
-
Comparative Study
Detection and classification of paraproteins by capillary immunofixation/subtraction.
A selection of 58 specimens with a monoclonal component identified by immunoelectrophoresis and/or immunofixation was analyzed with the immunosubtraction procedure on the Paragon 2000 capillary electrophoresis system. The capillary system detected 93% of the paraproteins and, using immunosubtraction, correctly identified 91% of the paraproteins. Paraproteins that were detected by immunofixation and/or immunoelectrophoresis but not by capillary electrophoresis were also missed by agarose electrophoresis and cellulose acetate electrophoresis. ⋯ Capillary electrophoresis and immunosubtraction correctly detected and identified three specimens containing biclonal paraproteins. In one specimen, capillary zone electrophoresis detected only one band, whereas agarose gel electrophoresis detected two bands. The sensitivity for detection and identification of oligoclonal banding by capillary electrophoresis was inferior to immunofixation.
-