Clinical chemistry
-
Randomized Controlled Trial Multicenter Study Clinical Trial
Clinical and analytical performance of the liaison cardiac troponin I assay in unstable coronary artery disease, and the impact of age on the definition of reference limits. A FRISC-II substudy.
Measurements of cardiac troponins are currently used as the standard for the detection of myocardial injury. None of the current assays complies with the new requirements on assay imprecision as proposed by the European Society of Cardiology/American College of Cardiology. Our aim was to evaluate the clinical and analytical performance of the Liaison cardiac troponin I (cTnI) assay. ⋯ The Liaison cTnI assay is a sensitive assay with a CV < or =10% at the 99th percentile URL. The ability to detect age-related differences among apparently healthy individuals is unique among today's commercial assays. The results indicate that different assays seem to identify different patient cohorts for cardiac risk in the lower range of cTnI concentrations.