Clinical chemistry
The most recent articles from:
Clin. Chem.
-
Accuracy and frequency of glucose measurement is essential to achieve safe and efficacious glucose control in the intensive care unit. Emerging continuous glucose monitors provide frequent measurements, trending information, and alarms. The objective of this study was to establish the level of accuracy of continuous glucose monitoring (CGM) associated with safe and efficacious glucose control in the intensive care unit. ⋯ The efficacy of CGM-informed and BG-informed commonly used glucose protocols is similar, but the risk of hypoglycemia may be reduced by use of CGM with MARD ≤ 10%. Protocol choice has greater influence on glucose control measures than the glucose measurement method.