Multicenter clinical and analytical evaluation of the AxSYM troponin-I immunoassay to assist in the diagnosis of myocardial infarction.
We evaluated the AxSYM troponin I (cTnI) immunoassay for assisting in the detection of acute myocardial infarction (AMI). At four sites, the total imprecision (CV) over 20 days was 6.3-10.2%. The minimum detectable concentration was 0.14 +/- 0.05 microgram/L. ⋯ This gave a diagnostic sensitivity of 91.8% and specificity of 92.4% when tested in serial samples collected within 24 h of admission in 633 patients presenting with chest pain, of which 122 had an AMI. The concordances of the AxSYM cTnI with the Stratus cTnI, OPUS cTnI, and Access cTnI were 95.3%, 95.1%, and 94.3%, respectively, from patients with suspected AMI. The AxSYM cTnI demonstrated excellent clinical specificity, >/=96%, in skeletal muscle injury, chronic renal disease, and same-day noncardiac surgery patients.
-
Multicenter Study
Portable blood gas and electrolyte analyzer evaluated in a multiinstitutional study.
A recently introduced blood gas/electrolyte analyzer (SenDx 100((R)), renamed ABL70) intended for point-of-care, near-patient, or stat laboratory use was evaluated simultaneously in four different institutions and compared with three different laboratory bench analyzers with respect to imprecision, inaccuracy (assessed by tonometry), and patient-sample analyses. The analyzer is equipped with a sensor cassette and a reagent cartridge for 50, 100, or 200 analyses and 100 or more traditional quality-control measurements. One analysis requires 170 microL of whole blood and takes <90 s. ⋯ However, the overall performance (in terms of CV and accuracy) was satisfactory in terms of clinical (e.g., CLIA '88) goals in all institutions. The mean difference and the CV of that difference in some 400 patient-sample comparisons were as follows: 0.010 (+/- 0.002%) for pH, -0.65 mmHg (+/- 4%) for PCO2, -0.49 mmHg (+/- 6%) for Po2, 0.44 mmol/L (+/- 1.2%) for sodium, -0.013 mmol/L (+/- 2.9%) for potassium, -0.016 mmol/L (+/- 2.6%) for ionized calcium, and -0.016 L/L (+/- 7. 1%) for the hematocrit. Its acceptable analytical performance and ease of operation make the SenDx 100 suitable for the analysis of blood gases and electrolytes.
-
Medicare will continue to increase its efforts to cut spending through aggressive review of claims and the use of new fraud and abuse regulations. Providers must be especially careful to provide correct procedure codes that define precisely what services have been provided and accurate diagnosis codes that link those procedures or tests to an appropriate diagnosis. ⋯ Coding and payment for new automated test panels are discussed, as well as the economic consequences of using smaller panels. Medicare coverage requirements, including medical necessity, are described, as well as the proper use of advance beneficiary notices and the Medicare appeals process.