Medical engineering & physics
-
Comparative Study
A low computational complexity algorithm for ECG signal compression.
In this work, a filter bank-based algorithm for electrocardiogram (ECG) signals compression is proposed. The new coder consists of three different stages. In the first one--the subband decomposition stage--we compare the performance of a nearly perfect reconstruction (N-PR) cosine-modulated filter bank with the wavelet packet (WP) technique. ⋯ The tests have been done for the 12 principal cardiac leads, and the amount of compression is evaluated by means of the mean number of bits per sample (MBPS) and the compression ratio (CR). The implementation cost for both the filter bank and the WP technique has also been studied. The results show that the N-PR cosine-modulated filter bank method outperforms the WP technique in both quality and efficiency.