Medical engineering & physics
-
Using a time-lapse image acquired from a CCD camera, we developed a non-contact and non-invasive device, which could measure both the respiratory and pulse rate simultaneously. The time-lapse image of a part of the subject's skin was consecutively captured, and the changes in the average image brightness of the region of interest (ROI) were measured for 30s. The brightness data were processed by a series of operations of interpolation as follows a first-order derivative, a low pass filter of 2 Hz, and a sixth-order auto-regressive (AR) spectral analysis. ⋯ Using AR spectral analysis, two clear peaks could be detected at approximately 0.3 and 1.2 Hz. The peaks were thought to correspond to the respiratory rate and the heart rate. Correlation coefficients of 0.90 and 0.93 were obtained for the measurement of heart rate and respiratory rate, respectively.