Medical engineering & physics
-
Flow diverters (FDs) are high density meshed stents designed to reduce blood flow into intra-cranial aneurysms. Though the FD is one of many intracranial aneurysm (IA) treatments, FD implantation may also result in the growth and rupture of residual aneurysms. The purpose of this research is to investigate the effect of FD implantation on IA hemodynamics. ⋯ The reason was that the flow resistance force at the aneurysm neck section was calculated to be lower than the flow driving force. In order to completely occlude the aneurysm, a higher resistance FD stent is to be required to suppress the dynamic forces. Patient-specific hemodynamic simulations offer means of quantitative estimation FD treatment outcomes.