Medical engineering & physics
-
Comparative Study
Biomechanical study of expandable pedicle screw fixation in severe osteoporotic bone comparing with conventional and cement-augmented pedicle screws.
Pedicle screws are widely utilized to treat the unstable thoracolumbar spine. The superior biomechanical strength of pedicle screws could increase fusion rates and provide accurate corrections of complex deformities. However, osteoporosis and revision cases of pedicle screw substantially reduce screw holding strength and cause loosening. ⋯ The pull-out strength was 5-fold greater than conventional pedicle screws and the failure energy was about 2-fold greater. Besides, the pull-out strength of expandable screw was reinforced by the expandable mechanism without cement augmentation, indicated that the risks of cement leakage from vertebral body would potentially be avoided. Comparing with the biomechanical performances of conventional screw with or without cement augmentation, the expandable screws are recommended to be applied for the osteoporotic vertebrae.