Comparative Study Clinical TrialMeasurements of gastric emptying of low- and high-nutrient liquids using 3D ultrasonography and scintigraphy in healthy subjects.
Scintigraphy represents the 'gold standard' for the measurement of gastric emptying. Recent studies suggest that three-dimensional (3D) ultrasonography may allow a precise measure of gastric emptying, given the capacity for accurate volume calculations of the stomach. The aim of this study was to compare measurements of gastric emptying of both low- and high-nutrient drinks by 3D ultrasonography with scintigraphy. ⋯ There was no significant difference between scintigraphic and ultrasonographic 50% emptying times (T50s) (soup: 27.7 +/- 4.8 min vs 23.8. +/- 4.8 min; dextrose: 122.2 +/- 13.3 min vs 131.9 +/- 10.2 min). There was a close correlation between scintigraphic and ultrasonographic T50s for both soup (r = 0.92, P = 0.0005) and dextrose (r = 0.88, P = 0.0007). For the T50s, the limits of agreement were -15.2 min and +8.1 min for the soup (mean difference -3.6 min) and -35.3 min and +47.6 min for dextrose (mean difference +6.2 min). 3D ultrasonography provides a valid measure of gastric emptying of liquid meals in healthy subjects.