To prospectively evaluate the perceived image quality of model-based iterative reconstruction (MBIR) compared to adaptive statistical iterative reconstruction (ASIR) and filtered back-projection (FBP) in computed tomography (CT) of the kidneys and retroperitoneum. ⋯ Compared to ASIR and FBP, MBIR provides superior perceived image quality, both overall and for several specific imaging features, across a broad range of tube current levels, and requires approximately half the radiation dose to achieve diagnostic overall perceived image quality. Accordingly, MBIR should enable CT scanning with improved perceived image quality and/or reduced radiation exposure.
-
To compare quality of ultra-low-dose thin-section computed tomography (CT) images of the lung reconstructed using model-based iterative reconstruction (MBIR) and adaptive statistical iterative reconstruction (ASIR) to filtered back projection (FBP) and to determine the minimum tube current-time product on MBIR images by comparing to standard-dose FBP images. ⋯ MBIR imaging shows higher overall quality with lower noise and streak artifacts than ASIR or FBP imaging, resulting in nearly 80% dose reduction without any degradations of overall image quality.
-
To evaluate knowledge-based iterative model reconstruction (IMR) to improve image quality and reduce radiation dose in coronary computed tomography angiography (cCTA). ⋯ IMR reduces intravascular noise on cCTA by 86%-88% compared to FBP, and improves image quality at radiation exposure levels 80% below our standard technique.