American journal of health-system pharmacy : AJHP : official journal of the American Society of Health-System Pharmacists
-
Am J Health Syst Pharm · Mar 1995
Comparative StudyEvaluation of flow rates for six disposable infusion devices.
The accuracy of flow was studied for six disposable infusion devices: Eclipse (Block Medical), Homepump (Block Medical), Intermate (Baxter Healthcare), MedFlo (Secure Medical), ReadyMED (McGaw), and SideKick (I-Flow). Each infusion device had a preset delivery rate of 100 mL/hr. Either 48 or 50 units of each device were filled with 100 mL of 0.9% sodium chloride injection. ⋯ The SideKick's mean infusion rate exceeded the +/- 15% range during the first half of the infusion. None of six disposable infusion devices delivered 100% of the volume to be infused within 60 minutes. Of the devices tested, the Homepump demonstrated the least variability in flow and the most complete volume delivery.