The flow rate accuracy and flow continuity of the Homepump, Infusor, Intermate, Paragon, Sidekick, and Springfusor 10 flow-restricting infusion devices were studied. Three of each device were tested in a temperature-controlled cabinet at the manufacturer-specified operating temperature and at 20 and 30 degrees C. The flow rates used were 100 ml/hr and the rate that would provide a 24-hour delivery of fluid, except in the case of the Springfusor 10, which was tested at the maximum and minimum flow rates. ⋯ The definition of the endopoint made little difference in the flow rate results for the Infusor and Springfusor 10 but significantly affected the flow rate results for the other devices, for which there was a substantial terminal phase with a prolonged period of reduced flow rate. The mean flow rate, error, and coefficient of variation for the Intermate devices stored under refrigeration were significantly different from those for the devices not refrigerated before use. When operated as recommended by the manufacturer, five of six devices had a flow rate error within the manufacturer's specifications; the results were affected by endpoint definition.