Clinical and applied thrombosis/hemostasis : official journal of the International Academy of Clinical and Applied Thrombosis/Hemostasis
-
Clin. Appl. Thromb. Hemost. · Jan 2017
Evaluation of a New Sonoclot Device for Heparin Management in Cardiac Surgery.
Sonoclot is used to measure kaolin-based activated clotting time (kACT) for heparin management. Apart from measuring kACT, the device assesses the patient's coagulation status by glass bead-activated tests (gbACTs; measuring also clot rate [CR] and platelet function [PF]). Recently, a new version of the Sonoclot has been released, and the redesign may result in performance changes. The aim of this study was to evaluate and compare the performance of the new (S2) and the previous (S1) Sonoclot. ⋯ Since clotting is faster in the new S2 compared to S1, shorter clotting times have to be considered in clinical practice. The use of S2 kACT in heparin management will result in higher heparin and protamine dosing unless heparin kACT target values are adjusted to correct for the differences in results between S1 and S2.
-
Clin. Appl. Thromb. Hemost. · Jan 2017
Thrombin Generation Assay in Hospitalized Nonsurgical Patients: A New Tool to Assess Venous Thromboembolism Risk?
Assessment of venous thromboembolism (VTE) risk is important to determine optimal primary prophylaxis in hospitalized patients. The Padua score helps to recognize patients with high VTE risk, but quantifying a VTE risk is often challenging in medical patients. Thrombin generation assay (TGA) reflects the pro-/anticoagulant balance and thus could help to better quantify VTE risk in medical hospitalized patients. ⋯ Single thrombin generation measurement could help to identify patients at risk of VTE in medical hospitalized patients.