Investigative radiology
-
Investigative radiology · Oct 2001
Individually adapted examination protocols for reduction of radiation exposure in chest CT.
To develop a simple directive for the reduction of radiation exposure without loss of diagnostic information in routine chest CT examinations. ⋯ By deriving mAs values from body weight estimation, an individually adapted protocol for chest CT can be recommended and easily employed in a clinical setting. With an adaptation of the tube current-time product based on the estimated body weight of the patient - 10 (body weight [kg] - 10 mAs), a well-balanced examination without significant loss of information, even in soft-tissue window settings, can be performed with this particular scanner. For this adapted mAs protocol, a mean reduction of radiation exposure of 45% was achievable, compared with the standard protocol. A maximum decrease per case down to 31 mAs was obtained, without relevant loss of image quality. Therefore, for other types of CT scanners, analogous protocols may be adapted.