Multiple-slice spin lock imaging of head and neck tumors at 0.1 Tesla: exploring appropriate imaging parameters with reference to T2-weighted spin-echo technique.
Spin lock imaging has been shown to be useful in characterizing head and neck tumors. The purposes of this study were to explore and develop multiple-slice spin lock gradient-echo (SL-GRE) sequences for head and neck imaging and to compare the tumor contrast on SL images to spin-echo (SE) T2-weighted images at 0.1 T. ⋯ The multiple-slice SL-GRE technique provides image contrast comparable to that of SE T2-weighted imaging for head and neck tumors at 0.1 T. With short locking pulse lengths and echo times, wide anatomic coverage and reduced motion and susceptibility artifacts can be achieved. The out-of-phase SL technique is useful in imaging salivary gland tumors.