Investigative radiology
-
Investigative radiology · Aug 2007
Comparative StudyBrain tumor enhancement in MR imaging at 3 Tesla: comparison of SNR and CNR gain using TSE and GRE techniques.
The purpose of this study was to compare brain and tumor signal characteristics of T1-weighted turbo spin-echo (TSE) and gradient recalled echo (GRE) sequence techniques at 3 T compared to TSE at 1.5 T, focusing on the detection of contrast enhancement, in a standardized animal model of a brain glioma. ⋯ With TSE at 3 T, the SNR gain comes close to the theoretically expected doubling with an even higher tumor CNR increase. In a clinical like setting at 3 T, where a T1w GRE sequence is used, tumor CNR gain is limited. Contrast dose should therefore not be decreased at 3 T.