Comparative StudyFeasibility of oscillometric aortic pressure and stiffness assessment using the VaSera VS-1500: comparison with a common tonometric method.
A number of operator-independent oscillometric devices to measure hemodynamics and arterial stiffness became available recently, but some and in particular VaSera VS-1500 do not provide estimates of aortic pressures and aortic pulse wave velocity (aPWV). The aim of this work was the retrospective application of the ARCSolver algorithm to pulse wave signals acquired with the VaSera VS-1500 device to estimate central systolic blood pressure (cSBP) and aPWV. ⋯ These results show that the combined application of the ARCSolver method with the VaSera VS-1500 device is feasible and the results are comparable with tonometric determination of cSBP and aPWV. This successful application of the ARCSolver may potentially help to improve cardiovascular risk stratification and prevention at an early stage in a community setting.