Blood pressure monitoring
-
Blood pressure monitoring · Feb 2018
Comparative StudyStudy of the disparity of noninvasive and invasive blood pressure measured by the Philips Intellivue MP50 monitor in surgeries inducing gall cardiac reflex under general anesthesia.
The aim of the present study was to comprehensively compare noninvasive and invasive blood pressure (BP) measured by the Philips Intellivue MP50 monitor in surgeries that may induce gall cardiac reflex under general anesthesia. ⋯ The oscillometric BP showed poor agreement with intra-arterial BP in cholecyst or bile duct surgeries that may induce gall cardiac reflex under general anesthesia. Therefore, according to the present data, application of oscillometric BP measured by the Philips Intellivue MP50 monitor in these surgery patients under general anesthesia cannot be recommended generally.
-
Blood pressure monitoring · Feb 2018
Validation of the Pangao PG-800A36 automatic wrist blood pressure monitor according to the European Society of Hypertension and the British Hypertension Society protocols.
The aim of this study was to validate the PG-800A36 automatic wrist blood pressure monitor according to the European Society of Hypertension International Protocol (ESH-IP) revision 2010 and the British Hypertension Society (BHS) protocols. ⋯ The Pangao PG-800A36 passed the requirements of the ESH-IP revision 2010 and achieved the A/A grade of the BHS protocol, which can be recommended for self-measurement in the general population.