Archives of disease in childhood
-
Clinical Trial
Home monitoring of transcutaneous oxygen tension in the early detection of hypoxaemia in infants and young children.
Twenty three patients (age range 0.5-40 months) with recurrent cyanotic episodes underwent physiological recordings, including transcutaneous oxygen tension (TcPO2) from a monitor modified for use at home (Kontron 821S). Of 69 episodes in which the arterial oxygen saturation (SaO2, Nellcor N200) was less than or equal to 80% for greater than or equal to 20 seconds and/or central cyanosis was present, the TcPO2 monitor alarmed (less than or equal to 20 mmHg or 2.67 kPa) in every episode. The pulse oximeter identified hypoxaemia in 62 out of 69 episodes, failing in seven episodes due to signal loss from movement artefact. ⋯ In 30 patients, the TcPo2 monitor also identified the gradual development of hypoxaemia, as confirmed by pulse oximetry. Twenty of these needed additional inspired oxygen and six subsequently needed ventilatory support in hospital. This TcPo2 monitor is a reliable detector of both sudden and gradual onset hypoxaemia and is able to be used by parents in the home.