The veterinary journal
-
The veterinary journal · Mar 2018
Comparative StudyBias, limits of agreement, and percent errors between acceleromyography and mechanomyography in anesthetized dogs.
This study compared measurements of neuromuscular function with mechanomyography (MMG) and acceleromyography (AMG) in nine anesthetized dogs receiving 0.1mg/kg vecuronium intravenously. Train-of-four (TOF) stimulation was applied to each ulnar nerve every 15s. The resulting amplitude of the first twitch (T1) and the TOF ratio were measured with both monitors. ⋯ Overall, T1 and TOF ratio had a small bias, wide LOA and PE>100%. Percent error was reduced to 30% when TOF≥0.7 was analyzed. Although there were no statistical differences between MMG and AMG in any variable of interest, individual discrepancies, wide LOA and high PE suggest that these monitors should not be used interchangeably for serial measurements on the same animals.