Neuromodulation : journal of the International Neuromodulation Society
-
The use of intraoperative neuromonitoring (IONM) has been adapted to address issues of safety and proper lead positioning in spinal cord stimulation. This multicenter case series seeks to incorporate the use of evoked compound action potential (ECAP) and late response (LR) recording and compare it with the results obtained with IONM, specifically electromyography (EMG), for the confirmation of lead placement. This study aimed to establish a correlation between ECAPs, LR, and EMG and publish human recordings of ECAPs and LR during their use with IONM. ⋯ LR correlated with EMG; leads with bilateral (not necessarily symmetric) EMG activity showed LR:ECAP > 1.5. An LR:ECAP of <1, with LR/EMG generated before the ECAP, indicated that the lead is too lateral. The use of ECAP and LR has the potential of maintaining objective lead placement, without the need for needle placement with IONM.