Journal of clinical monitoring and computing
-
J Clin Monit Comput · Apr 2005
Cerebral monitoring in the operating room and the intensive care unit - an introductory for the clinician and a guide for the novice wanting to open a window to the brain. Part II: Sensory-evoked potentials (SSEP, AEP, VEP).
An evoked potential differs from the EEG mainly in two ways: 1. The EEG is a random, continuous signal, which arises from the ongoing activity of the outer layers of the cortex. An evoked potential is the brain's response to a repetitive stimulus along a specific nerve pathway. 2. ⋯ Evoked potentials are used both as a diagnostic tool and as a monitoring technique. As diagnostic tests, evoked potentials are useful to evaluate neurologic disorders such as: a) multiple sclerosis, b) acoustic nerve tumors, and c) optic neuritis. As a monitoring modality, evoked potentials are used during all surgical procedures, which might compromise part of the brain or the spinal cord.
-
J Clin Monit Comput · Apr 2005
Cerebral monitoring in the operating room and the intensive care unit: an introductory for the clinician and a guide for the novice wanting to open a window to the brain. Part I: The electroencephalogram.
While there is an increasing body of knowledge in regard to central nervous system function and/or the mode of action of centrally active agents on neuronal function, little is done to develop new techniques on how to measure such changes. Also, monitoring of the cardiovascular system in the past has made extensive progress especially when it comes to evaluate the failing heart. In contrast monitoring of the central nervous system is only done in rare cases where operative procedures likely impede nervous function integrity. ⋯ Gives suggestions for assessing and improving signal quality, including noise and artifact rejection, which usually are encountered in the operation room and the intensive care unit, both of which can be considered as electrically contaminated. Gives examples of EEG power spectra and evoked potential monitoring related to different types of anesthesia, in coma, after head trauma, and for the detection of ischemic events. In addition, gives complete coverage of those machines being available for the OR and the ICU, including a list of parameters regarding latency and amplitude in evoked potential As an introductory, recommendations are given for the novice to start cerebral monitoring and guide the beginner in setting up cerebral monitoring in the clinical environment.