Journal of clinical monitoring and computing
-
J Clin Monit Comput · Dec 2012
Multicenter StudyThe ability of a new continuous cardiac output monitor to measure trends in cardiac output following implementation of a patient information calibration and an automated exclusion algorithm.
A new non-invasive continuous cardiac output (esCCO) monitoring system solely utilizing a routine cardiovascular monitor was developed, even though a reference cardiac output (CO) is consistently required. Subsequently, a non-invasive patient information CO calibration together with a new automated exclusion algorithm was implemented in the esCCO system. We evaluated the accuracy and trending ability of the new esCCO system. ⋯ The percentage error was 69.6 %. Polar plots analysis showed that the mean polar angle was -1.6° and radial limits of agreement were ±53.3°. This study demonstrates that the patient information calibration is clinically useful as ICO, but trending ability of the new esCCO system is not clinically acceptable as judged by percentage error and polar plots analysis, even though it's trending ability is comparable with currently available arterial waveform analysis methods.