Journal of clinical monitoring and computing
-
J Clin Monit Comput · Oct 2019
Multicenter Study Comparative Study Clinical Trial Observational StudyMindray 3-directional NMT Module (a new generation "Tri-axial" neuromuscular monitor) versus the Relaxometer mechanomyograph and versus the TOF-Watch SX acceleromyograph.
Recently introduced Mindray "3-directional" neuromuscular transmission transducer (NMT, Shenzhen, China) acceleromyograph) claim to monitor thumb movement in 3 different directions. We compared NMT with the gold standard Relaxometer® mechanomyograph (MMG, Groningen University, Netherlands) in Study-1 and with TOF-Watch SX™ (WTCH) acceleromyograph from which it was developed in Study-2. We used first twitch (T1%) and train-of-four (TOF) ratio rocuronium 0.6 mg kg-1 neuromuscular block to evaluate NMT diagnostic accuracy in indicating 3 clinically relevant time points namely; MMG T1 5% (95% twitch depression) for tracheal intubation, MMG T1 25% for repeat neuromuscular blocking agents (NMBAs) administration, and MMG 0.9 TOF ratio full neuromuscular block recovery. ⋯ NMT could not efficaciously detect MMG time for tracheal intubation; NMBAs repeat dose administration or full neuromuscular block recovery. Data from NMT cannot be used interchangeably with MMG. Our study revealed that NMT Tri-axial acceleromyography seems to offer no advantage over the MMG gold standard or the classic Mono-axial TOF-Watch SX monitor.