Journal of science and medicine in sport
-
Determination of total haemoglobin mass (Hbmass) via carbon monoxide (CO) depends critically on repeatable measurement of percent carboxyhaemoglobin (%HbCO) in blood with a hemoximeter. The main aim of this study was to determine, for an OSM3 hemoximeter, the number of replicate measures as well as the theoretical change in percent carboxyhaemoglobin required to yield a random error of analysis (Analyser Error) of ≤1%. Before and after inhalation of CO, nine participants provided a total of 576 blood samples that were each analysed five times for percent carboxyhaemoglobin on one of three OSM3 hemoximeters; with approximately one-third of blood samples analysed on each OSM3. ⋯ For duplicate analyses of blood, the Analyser Error for the 2-min method was 3.7, 4.0 and 5.0% for the three OSM3s when the percent carboxyhaemoglobin increased by two above resting values. With quintuplicate analyses of blood, the corresponding errors reduced to .8, .9 and 1.0% for the 2-min method when the percent carboxyhaemoglobin increased by 5.5 above resting values. In summary, to minimise the Analyser Error to ∼≤1% on an OSM3 hemoximeter, researchers should make ≥5 replicates of percent carboxyhaemoglobin and the volume of CO administered should be sufficient increase percent carboxyhaemoglobin by ≥5.5 above baseline levels.