J Appl Clin Med Phys
-
J Appl Clin Med Phys · Jan 2019
Improvements in treatment planning calculations motivated by tightening IMRT QA tolerances.
Implementing tighter intensity modulated radiation therapy (IMRT) quality assurance (QA) tolerances initially resulted in high numbers of marginal or failing QA results and motivated a number of improvements to our calculational processes. This work details those improvements and their effect on results. One hundred eighty IMRT plans analyzed previously were collected and new gamma criteria were applied and compared to the original results. ⋯ After investigating and improving several elements of the IMRT calculation processes, the frequency of questionable plans was reduced to 11% and that of failing plans to less than 1%. Tighter IMRT QA tolerances revealed the need to improve several elements of our plan calculations. As a consequence, the accuracy of our plans have improved, and the frequency of finding marginal or failing IMRT QA results, remains within our practical ability to respond.