Comparative Study
The effects of different correction techniques on absolute volume determination with SPECT using a threshold edge detection method.
Quantitation of planar radionuclide images is hampered by structures containing radioactivity which overlie or underlie the organ of interest. The introduction of single photon emission computerized tomography (SPECT) overcame this problem to a large extent and enhanced the contrast of the images. Attenuation of photons, however, degrades the resultant SPECT images and correction methods for photon absorption and scatter were subsequently proposed. ⋯ A negative correlation was found between threshold and volume. This reduction in threshold was most prominent when scatter and attenuation correction were combined. This study shows that correction methods for attenuation of photons influence the threshold value for volume quantitation and the use of a constant threshold value could lead to underestimation of larger volumes.