Permanent transperineal interstitial 125I and 103Pd prostate implants are generally planned to deliver a specific dose to a clinically defined target volume; however, the post-implant evaluation usually reveals that the implant delivered a lower or higher dose than planned. This difference is generally attributed to such factors as source placement errors, overestimation of the prostate volume on CT, and post-implant edema. In the present work we investigate the impact of edema alone. ⋯ A comparison of the DVH of the static model to the 30 edema corrected DVHs revealed that the plan overestimated the total dose by an amount that increased with the magnitude of the edema and the edema half-life. The maximum overestimation was 15% for 125I and 32% for 103Pd. For more typical edema parameters (a 50% increase in volume and a 10 day half-life) the static plan for 125I overestimated the total dose by about 5%, whereas that for 103Pd overestimated it by about 12%.