Med Phys
-
Subsecond spiral computed tomography (CT) offers great potential for improving heart imaging. The new multi-row detector technology adds significantly to this potential. We therefore developed and validated dedicated cardiac reconstruction algorithms for imaging the heart with subsecond multi-slice spiral CT utilizing electrocardiogram (ECG) information. ⋯ Moreover, if technically feasible, lower rotation times such as 0.3 s or even less would result in improved image quality. The use of multi-slice techniques for cardiac CT together with the new z-interpolation methods improves the quality of heart imaging significantly. The high temporal resolution of 180 degrees MCI is adequate for spatial and temporal tracking of anatomic structures of the heart (4D reconstruction).