Med Phys
-
Comparative Study Clinical Trial
Segmentation of IMRT plans for radical lung radiotherapy delivery with the step-and-shoot technique.
The purpose of this work was to determine a segmentation protocol for the treatment of localized non-small-cell lung cancer (NSCLC) with intensity-modulated radiotherapy (IMRT) that is as effective as possible while practically simple and hence robust to known practical inaccuracies. This study focused on the stratification of continuous profiles into a discrete number of intensity levels. The selection of the segmentation parameters for the delivery of the fluence profiles using multiple static fields has been considered. ⋯ All segmented IMRT plans gave a better lung sparing than the conformal plans, indicating that the deterioration of IMRT plans following segmentation is not equivalent to that of unmodulated, conformal plans. However, optimized conformal plans have the potential to approach the lung sparing achieved by segmented IMRT plans. Among the IMRT situations examined, five-field treatment plans for the lung, utilizing a maximum of 40 segments in total, have proven to give a good approximation of the IMRT plans with continuous modulation.