Med Phys
-
Two reasons for the recent rise in radiation exposure from CT are increases in its clinical applicability and the desire to maintain high SNR while acquiring smaller voxels. To address this emerging dose problem, several strategies for reducing patient exposure have already been proposed. One method employed in cardiac imaging is ECG-driven modulation of the tube current between 100% at one time point in the cardiac cycle and a reduced fraction at the remaining phases. ⋯ Our approach was validated using various data obtained with simulated and scanner-implemented dose modulation. We determined that as little as 10% of the total dose was required to reproduce full quality images with a 1 mm spatial error and an error in intensity values on the order of the image noise. Thus, our technique offers considerable dose reductions compared to standard imaging protocols, with minimal effects on the quality of the final data.