Med Phys
-
The variances and biases inherent in quantifying PET tracer uptake from instrumentation factors are needed to ascertain the significance of any measured differences such as in quantifying response to therapy. The authors studied the repeatability and reproducibility of serial PET measures of activity as a function of object size, acquisition, reconstruction, and analysis method on one scanner and at three PET centers using a single protocol with long half-life phantoms. ⋯ Short-term scanner variability is low compared to other sources of error. There are tradeoffs in noise and bias depending on acquisition, processing, and analysis methods. The SD of a new PET measure can be estimated from a known SD if the ratios of available coincident counts between the two PET scanner acquisitions are known and both employ the same ROI definition. Results suggest it is feasible to use PET/CTs from different vendors and sites in clinical trials if they are properly cross-calibrated.