Med Phys
-
To calculate imaging doses to the rectum, bladder, and femoral heads as part of a prostate cancer treatment plans, assuming an image guided radiation therapy (IGRT) procedure involving either the multidetector CT (MDCT) or kilovoltage cone-beam CT (kV CBCT). ⋯ IGRT procedures can irradiate the OARs to an imaging dose level that is great enough to require careful evaluation and perhaps even adjustment of original treatment planning in order to still satisfy the dose constraints. This study only considered one patient CT because the CT x rays cover a relatively larger volume of the body and the dose distribution is considerably more uniform than those associated with the therapeutic beams. As a result, the dose to an organ from CT imaging doses does not vary much from one patient to the other for the same CT settings. One factor that would potentially affect such CT dose level is the size of the patient body. More studies are needed to develop accurate and convenient methods of accounting for the imaging doses as part of treatment planning.