Med Phys
-
Previously, the diagnostic radiological index of protection (DRIP) was proposed as a metric for quantifying the protective value of radioprotective garments. The DRIP is a weighted sum of the percent transmissions of different radiation beams through a garment. Ideally, the beams would represent the anticipated stray radiation encountered during clinical use. However, it is impractical to expect a medical physicist to possess the equipment necessary to accurately measure transmission of scattered radiation. Therefore, as a proof of concept, the authors tested a method that applied the DRIP to clinical practice. ⋯ A method for measuring the DRIP for protective garments using scatter-mimicking primary beams was developed. There was little discernable advantage in protective value per unit weight for lead-alternative versus lead-only garments. Careful consideration must be given to the balance of protection and weight when choosing a lead-alternative protective garment with a lower specified "lead equivalence," e.g., 0.35 mm. The DRIP has the potential to resolve this dilemma. Reporting the DRIP relative to areal density is an ideal metric for objective comparisons of protective garment performance, considering both protective value in terms of transmission of radiation and garment weight.