Diabetes
-
The stability and longevity of the polyethylene-polypropylene glycol-stabilized insulin have been tested in vitro and in vivo in an implanted insulin-infusion device, the programmable implantable medication system (PIMS). In vitro tests demonstrated long-term compatibility with refill cycles of up to 3 mo, with a preparation of 400 U/ml. Total test period in vitro was 3.2 device-yr (combined time of device use). ⋯ There has been no insulin blockage of the catheter of active pumps after 5.1 dog-yr (combined time of trials) of trials (up to 5 mo between refills in a single dog). Structural stability of insulin was analyzed by high-performance liquid chromatography. On average, 90.8% of the insulin sampled from the reservoir in vivo was native insulin, with an average of 96.2% retention of active forms.