Adv Exp Med Biol
-
Comparative Study
Pulse oximetry and transcutaneous oxygen tension for detection of hypoxemia in critically ill infants and children.
We tested the performance of transcutaneous oxygen monitoring (TcPO2) and pulse oximetry (tcSaO2) in detecting hypoxia in critically ill neonatal and pediatric patients. In 54 patients (178 data sets) with a mean age of 2.4 years (range 1 to 19 years), arterial saturation (SaO2) ranged from 9.5 to 100%, and arterial oxygen tension (PaO2) from 16.4 to 128 mmHg. Linear correlation analysis of pulse oximetry vs measured SaO2 revealed an r value of 0.95 (p less than 0.001) with an equation of y = 21.1 + 0.749x, while PaO2 vs tcPO2 showed a correlation coefficient of r = 0.95 (p less than 0.001) with an equation of y = -1.04 + 0.876x. ⋯ Pulse oximetry was reliable at values above 65%, but was inaccurate and overestimated the arterial SaO2 at lower values. TcPO2 tended to underestimate the arterial value with increasing PaO2. Pulse oximetry had the best sensitivity to specificity ratio for hypoxia between 65 and 90% SaO2; for tcPO2 the best results were obtained between 35 and 55 mmHg PaO2.
-
The integration of an ECG-electrode into a common transcutaneous sensor allows simple handling and leads to a reduction of the physiological stress of pre-term infants. Furthermore it may allow future replacement of an invasive method to measure the ECG under labour by a non-invasive one.