American journal of veterinary research
-
To compare results reported for blood gas partial pressures, electrolyte concentrations, and Hct in venous blood samples collected from cattle, horses, and sheep and analyzed by use of a portable clinical analyzer (PCA) and reference analyzer (RA). ⋯ Data from blood gas and electrolyte analyses obtained by use of the PCA can be used to evaluate the health status of cattle, horses, and sheep. Furthermore, the handheld PCA device may have a great advantage over the RA device as a result of the ability to analyze blood samples on farms that may be located far from urban centers.