Ultrasonics
-
The objective of this study was to assess in vitro the impact of ultrasound scanner settings and contrast bolus volume on time-intensity curves formed from dynamic contrast-enhanced ultrasound image loops. An indicator-dilution experiment was developed with an in vitro flow phantom setup used with SonoVue contrast agent (Bracco SpA, Milan, Italy). Imaging was performed with a Philips iU22 scanner and two transducers (L9-3 linear and C5-1 curvilinear). ⋯ One should be aware of these issues to standardize their technique in each specific organ of interest and to achieve accurate, sensitive, and reproducible data using dynamic contrast-enhanced ultrasound. One way to mitigate the impact of ultrasound scanner settings in longitudinal, multi-center quantitative dynamic contrast-enhanced ultrasound studies may be to prohibit any adjustments to those settings throughout a given study. Further clinical studies are warranted to confirm the reproducibility and diagnostic or prognostic value of time-intensity curve parameters measurements in a particular clinical scenario of interest, for example that of cancer patients undergoing vascular targeting therapies.