Neurosurg Focus
-
This study was conducted to document the extent to which flow depends on valve position in relation to head-pressure reference. Medtronic PS Medical Delta valves (contour model, performance levels 0.5, 1.0, 1.5, and 2.0) were studied in a bench test designed to evaluate flow rates with respect to valve position in relation to the head-pressure reference postion. The valves were connected to an "infinite" reservoir by the standard inlet catheter. ⋯ Movement of the valve above this level results in a net decrease in effective head pressure, and movement below this position results in a net increase in effective head pressure. Thus, the positioning of shunt valves in locations different from this pressure reference position should be performed only with the knowledge that significant increases in outflow rate may occur when the valve body is positioned lower than the inlet catheter tip. This increase in outflow rate is not the result of siphoning or a defect in the antisiphon device but instead the result of a net increase in effective head pressure.